Z-1034-2022 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1034-2022 was initiated by Olympus Corporation Of The Americas on Mar 18, 2022 and is designated Class II. Reason for recall: GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients The recall status is ongoing. Affected quantity: 731 units.

Recall Details

Product Type
Devices
Report Date
May 11, 2022
Initiation Date
Mar 18, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
731 units

Product Description

Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.

Reason for Recall

GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients

Distribution Pattern

Nationwide

Code Information

All serial numbers. GTIN: 04953170356322

Source