Z-1046-2022 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1046-2022 was initiated by Olympus Corporation Of The Americas on Mar 8, 2022 and is designated Class II. Reason for recall: Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a… The recall status is ongoing. Affected quantity: 60 units.

Recall Details

Product Type
Devices
Report Date
May 18, 2022
Initiation Date
Mar 8, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
60 units

Product Description

BF-MP60: OES Bronchofiberscope

Reason for Recall

Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using sterilization

Distribution Pattern

US Nationwide distribution.

Code Information

All serial numbers. UDI: 04953170339394