Z-1048-2022 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

Recall Details

Product Type
Devices
Report Date
May 18, 2022
Initiation Date
March 8, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
292 units

Product Description

BF-MP160F: EVIS EXERA Bronchofibervideoscope

Reason for Recall

Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using sterilization

Distribution Pattern

US Nationwide distribution.

Code Information

All serial numbers. UDI: 04953170289064