Z-1050-2015 Class II Terminated
FDA device recall Z-1050-2015 was initiated by Hospira Inc. on Jun 11, 2014 and is designated Class II. Reason for recall: Plum A+ infusion system pole clamp assemblies were discovered to be cracked. If the knob on the pole clamp assembly is cracked, the infuser cannot be attached to the IV pole since the knob is used to… The recall status is terminated (terminated Nov 28, 2016). Affected quantity: 102 infusors.
Recall Details
- Product Type
- Devices
- Report Date
- Feb 11, 2015
- Initiation Date
- Jun 11, 2014
- Termination Date
- Nov 28, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 102 infusors
Product Description
Plum A+ Infusion System with Hospira MedNet List 20792-04-25, Pole Clamp Assembly Part Number 820-95443-008. A cassette based multi-function infusion system. Each system includes a pumping module and an assortment of disposable IV sets, optional accessories, and an operator's manual. The pole clamp assembly is used to provide a mechanism to tighten and loosen the grip of the pole clamp to remove the infuser from the IV pole.
Reason for Recall
Plum A+ infusion system pole clamp assemblies were discovered to be cracked. If the knob on the pole clamp assembly is cracked, the infuser cannot be attached to the IV pole since the knob is used to tighten and loosen the grip of the pole clamp to the pole.
Distribution Pattern
US Distribution to New York.
Code Information
List Number: 20792-04-25; Serial Numbers: 17737372 ; 17737375 ; 17737377 ; 17737378 ; 17737383 ; 17737385 ; 17737390 ; 17737394 ; 17737395 ; 17737396 ; 17737397 ; 17737398 ; 17737399 ; 17737400 ; 17737402 ; 17737405 ; 17737408 ; 17737409 ; 17737414 ; 17737416 ; 17737419 ; 17737420 ; 17737421 ; 17737422 ; 17737424 ; 17737425 ; 17737426 ; 17737429 ; 17737430 ; 17737431 ; 17737432 ; 17737433 ; 17737435 ; 17737438 ; 17737440 ; 17737442 ; 17737443 ; 17737444 ; 17737445 ; 17737446 ; 17737447 ; 17737448 ; 17737449 ; 17737450 ; 17737451 ; 17737452 ; 17737454 ; 17737456 ; 17737458 ; 17737459 ; 17737460 ; 17737462 ; 17737463 ; 17737464 ; 17737467 ; 17737470 ; 17737475 ; 17737476 ; 17737478 ; 17737479 ; 17737480 ; 17737483 ; 17737485 ; 17737486 ; 17737489 ; 17737490 ; 17737492 ; 17737498 ; 17737500 ; 17737501 ; 17737502 ; 17737506 ; 17737507 ; 17737508 ; 17737511 ; 17737514 ; 17737516 ; 17737518 ; 17737520 ; 17737521 ; 17737522 ; 17737524 ; 17737525 ; 17737526 ; 17737531 ; 17737533 ; 17737535 ; 17737539 ; 17737542 ; 17737545 ; 17737546 ; 17737550 ; 17737551 ; 17737556 ; 17737557 ; 17737558 ; 17737559 ; 17737560 ; 17737562 ; 17737564 ; 17737569 ; 17737575
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1050-2015
- openFDA file date
- Oct 27, 2026
- Loaded by RegDataLab