Z-1062-2024 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- March 6, 2024
- Initiation Date
- January 10, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,070 units
Product Description
AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F
Reason for Recall
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
Distribution Pattern
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Code Information
UDI/DI Case: 50190752114188, Each: 10190752114180; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.