Z-1061-2024 Class I Ongoing

Recalled by Vyaire Medical — Mettawa, IL

FDA device recall Z-1061-2024 was initiated by Vyaire Medical on Jan 10, 2024 and is designated Class I. Reason for recall: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no… The recall status is ongoing. Affected quantity: 3,683,407 units.

Recall details

Product Type
Devices
Report Date
Mar 6, 2024
Initiation Date
Jan 10, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,683,407 units

Product description

AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8005

Reason for recall

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

Distribution pattern

Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa

Code information

UDI/DI Case: 50190752114157, Each: 10190752114159; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1061-2024
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Vyaire Medical? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Vyaire Medical: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Vyaire Medical →

Source: openFDA record for Z-1061-2024. Report a data issue: hello@regdatalab.com.