Z-1065-2024 Class I Ongoing
FDA device recall Z-1065-2024 was initiated by Vyaire Medical on Jan 10, 2024 and is designated Class I. Reason for recall: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no… The recall status is ongoing. Affected quantity: 1,800,000 units.
Recall details
- Product Type
- Devices
- Report Date
- Mar 6, 2024
- Initiation Date
- Jan 10, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,800,000 units
Product description
Reason for recall
Distribution pattern
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1065-2024
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026