Z-1077-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 17, 2021
- Initiation Date
- December 10, 2020
- Termination Date
- January 31, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 583 units
Product Description
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18N, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, Sterile EO,
Reason for Recall
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Distribution Pattern
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
Code Information
Lot Number: 1394204, 1424962, 1464709, 1473181, 1477229, 1481735, 1495182