Z-3192-2024 Class II Ongoing

Recalled by Argon Medical Devices, Inc — Athens, TX

FDA device recall Z-3192-2024 was initiated by Argon Medical Devices, Inc on Jul 2, 2024 and is designated Class II. Reason for recall: Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site… The recall status is ongoing. Affected quantity: 95,430.

Recall details

Product Type
Devices
Report Date
Oct 2, 2024
Initiation Date
Jul 2, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
95,430

Product description

BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370-1080-03, 370-1580-01, 370-1580-02, 370-1580-03; Tru-Core II Biopsy Instrument, REF: 763114100x, 763116100X, 763116160X, 763118100X, 763118200X, 763120100X, 763120160X, 763418200X, 763418250X, 763114200X

Reason for recall

Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.

Distribution pattern

Worldwide - US Nationwide distribution including in the states of AZ, FL, OH, NY, MO, CO, IL, NC, CA, AK, PA, MN, ND, TX, SC, AR, WI, OK, TN, KS, UT, NJ, VA, MA, KY, IN, AL, DE, MT, NE, WV, NV, HI, OR, WA, NH, LA, CT, MI, GA, MD, ID, MS, IA, RI, NM, VT and the countries of ITA, TUN, ROU, LTU, BEL, GBR, CAN, TUR, POL, PAN, THA, PRT, TWN, AUT, NZL, AUS, BGR, MEX, CHE, NLD, PHL, PER, FRA, SGP, ZAF, CZE, EST, ESP, OMN, NPL, COL, PRI, ISR, JPN, KEN, BHR, ARE, CHN.

Code information

REF/Box UDI-DI/Unit UDI-DI/Lot(Expiration Date): 360-1080-01/20886333004373/00886333004379/11563436(22-Mar-2027), 11568010(16-Apr-2027), 11568790(1-May-2027), 11569552(7-May-2027), 11572560(20-May-2027) 360-1080-02/20886333224009/00886333224005/11563678(22-Mar-27), 11562475(05-Mar-27), 11562476(14-Mar-27), 11566932(08-Apr-27), 11567139(11-Apr-27), 11568361(16-Apr-27), 11569013(23-Apr-27), 11569844(29-Apr-27), 11571094(02-May-27), 11572044(08-May-27), 11572886(26-May-27); 360-1080-03/20886333224016/00886333224012/11569845(22-Apr-27), 11562477(04-Mar-27), 11564861(26-Mar-27), 11567888(12-Apr-27), 11568554(18-Apr-27), 11570075(04-May-27); 360-1580-01/20886333004359/00886333004355/11563438(21-Mar-2027), 11563439(19-Mar-2027); 360-1580-02/20886333224023/00886333224029/11562478(14-Mar-27), 11563440(18-Mar-27), 11563682(20-Mar-27), 11565238(03-Apr-27), 11566882(09-Apr-27), 11570132(01-May-27), 11570814(09-May-27), 11573339(20-May-27); 360-1580-03/20886333224030/00886333224036/11570288(02-May-27), 11563442(20-Mar-27), 11566173(03-Apr-27), 11566883(10-Apr-27), 11568153(16-Apr-27), 11570813(08-May-27); 360-2080-01/20886333004342/00886333004348/11563564(26-Mar-2027), 11566036(8-Apr-2027), 11567992(16-Apr-2027), 11570142(1-May-27), 11570863(8-May-27); 360-2080-02/20886333224047/00886333224043/11562964(07-Mar-27), 11563922(15-Mar-27), 11564189(22-Mar-27); 360-2080-03/20886333224054/00886333224050/11563443(15-Mar-27), 11566936(11-Apr-27), 11567889(22-Apr-2027); 370-1080-01/20886333004380/00886333004386/11564795(26-Mar-27); 370-1080-02/20886333224061/00886333224067/11562608(05-Mar-27), 11562748(04-Mar-27), 11564843(26-Mar-27), 11566037(08-Apr-27), 11571101(30-Apr-27), 11571360(06-May-27); 370-1080-03/20886333224078/00886333224074/11564844(22-Mar-27), 11566690(16-Apr-27), 11571100(01-May-27), 11573030(13-May-27); 370-1580-01/20886333004366/00886333004382/11564366(26-Mar-27); 370-1580-02/20886333224085/00886333224081/11571824(06-May-27), 11562114(07-Mar-27), 11566038(08-Apr-27), 11571824(06-May-27); 370-1580-03/20886333224092/00886333224098/11569414(22-Apr-27), 11571104(01-May-27); 763114100x/20886333006643/00886333006649/11563673(11-May-2029), 11570356(23-Apr-2029), 11571058(4-Apr-2029), 11572456(29-May-2029), 11572868(13-May-2029), 11573300(22-May-2029); 763116100X/20886333006612/00886333006618/11562815(7-Mar-2029), 11571061(29-Apr-2029); 763116160X/20886333006605/00886333006601/11571063(1-May-2029); 763118100X/20886333006582/00886333006588/11564330(26-Mar-2029); 763118200X/20886333006650/00886333006656/11562085(5-Mar-2029), 11571065(4-APR-2029); 763120100X/20886333006568/00886333006564/11564331(26-Mar-2029); 763120160X/20886333006575/00886333006571/11565618(10-Apr-2029); 763418200X/20886333006803/00886333006809/11566213(10-Apr-2029); 763418250X/20886333006810/00886333006816/11564594(22-Mar-2029), 11565619(4-Apr-2029), 11572175(22-May-2029); 763114200X/20886333006629/00886333006625/11566106(4-Apr-2029)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3192-2024
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Argon Medical Devices, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-3192-2024. Report a data issue: hello@regdatalab.com.