Z-0563-2022 Class II Ongoing
FDA device recall Z-0563-2022 was initiated by Argon Medical Devices, Inc on Jan 24, 2022 and is designated Class II. Reason for recall: Guidewire included with two lots of Arterial Line Kits would not fit through the needle of the same kit. This could potentially lead to a slight delay in the procedure and minor blood loss. The recall status is ongoing. Affected quantity: 600 ea.
Recall details
- Product Type
- Devices
- Report Date
- Feb 16, 2022
- Initiation Date
- Jan 24, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 600 ea
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the states of AL, CA, DE, IL, IN, LA, MT, NV, OH, TX, VA, TN.