Z-0563-2022 Class II Ongoing

Recalled by Argon Medical Devices, Inc — Athens, TX

Recall Details

Product Type
Devices
Report Date
February 16, 2022
Initiation Date
January 24, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
600 ea

Product Description

1. Arterial Line Kit 20ga x 6" 2. Arterial Line Kit /1 20ga x 6"

Reason for Recall

Guidewire included with two lots of Arterial Line Kits would not fit through the needle of the same kit. This could potentially lead to a slight delay in the procedure and minor blood loss.

Distribution Pattern

US Nationwide distribution in the states of AL, CA, DE, IL, IN, LA, MT, NV, OH, TX, VA, TN.

Code Information

1. Model: 400115A and Serial # 11377499, Expire Date: 10/27/2022 2. Model: 498100 and Serial # 11378381, Expire Date: 4/15/2026