Z-1099-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 17, 2019
- Initiation Date
- September 14, 2018
- Termination Date
- September 14, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6 devices
Product Description
SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.
Reason for Recall
For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).
Distribution Pattern
US Distribution to states of: IN, MO, and OH.
Code Information
Serial numbers 16S13187, 16S14887, 16S15335, 16S15336, 16S15337, and 16S11877.