Z-1113-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 25, 2022
- Initiation Date
- March 25, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1575 units total: 1183 US; 392 OUS
Product Description
VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198
Reason for Recall
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.
Distribution Pattern
Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia Singapore Spain Sweden United Kingdom
Code Information
Until further notice, future lots will also be affected. UDI: 10758750008902 Affected Lot Number/ Expiration Date: 1910 14-AUG-2022 1920 19-SEP-2022 1930 19-OCT-2022 1940 31-OCT-2022 1950 28-JAN-2023 1960 31-JAN-2023