Z-1113-2022 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, Inc — Raritan, NJ

FDA device recall Z-1113-2022 was initiated by Ortho-Clinical Diagnostics, Inc on Mar 25, 2022 and is designated Class II. Reason for recall: Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and… The recall status is ongoing. Affected quantity: 1575 units total: 1183 US; 392 OUS.

Recall Details

Product Type
Devices
Report Date
May 25, 2022
Initiation Date
Mar 25, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1575 units total: 1183 US; 392 OUS

Product Description

VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198

Reason for Recall

Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.

Distribution Pattern

Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia Singapore Spain Sweden United Kingdom

Code Information

Until further notice, future lots will also be affected. UDI: 10758750008902 Affected Lot Number/ Expiration Date: 1910 14-AUG-2022 1920 19-SEP-2022 1930 19-OCT-2022 1940 31-OCT-2022 1950 28-JAN-2023 1960 31-JAN-2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1113-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.