Z-1123-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 17, 2019
- Initiation Date
- September 21, 2018
- Termination Date
- September 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 22
Product Description
Exprt Precision System: Revision Hip, Proximal Body with Bolt, Lateral Offset, Sterile R, djo surgical, REF: 495-00-065, 495-00-075, 495-00-085, 495-01-065, 495-01-075, 495-01-085
Reason for Recall
Complaints regarding loose proximal body bolts or bolts that were found through x-ray as sitting above the proximal body and not properly seated. Patient risks may include: pain, dissociation, and possible revision surgery
Distribution Pattern
US: TN, MN, TX, CO, IL, IN, OH, SC, MS, MI, RI, PA, ME, NY, WA, MO, AR, OK, LA, AL, FL, GA, UT, AZ, CA, NV,.
Code Information
All lots.