Z-1234-2026 Class II Ongoing

Recalled by Encore Medical, LP — Austin, TX

Recall Details

Product Type
Devices
Report Date
February 11, 2026
Initiation Date
January 5, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
49

Product Description

EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded, HXe+, 40G, REF: 942-01-40G; Liner, 10 degree Hooded, HXe+, 32C, REF: 942-01-32C

Reason for Recall

942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.

Distribution Pattern

US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CA

Code Information

REF/UDI-DI/Lot(Expiration): 942-01-36H/00190446309789/624Z1032(03-15-2026), 942-01-36J/00190446309802/626Z1021(03-15-2026), 942-01-40G/00190446309826/628Z1034(03-15-2026), 942-01-32C/00190446309697/615Z1026(03-15-2026)