Z-1160-2013 Class I Terminated
FDA device recall Z-1160-2013 was initiated by Hospira Inc. on Mar 18, 2013 and is designated Class I. Reason for recall: Battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used. The recall status is terminated (terminated Mar 7, 2017). Affected quantity: List Number 13086: 286 pumps; List Number 13087: 45,376 pumps; List Number 13088: 23,492 pumps.
Recall Details
- Product Type
- Devices
- Report Date
- May 8, 2013
- Initiation Date
- Mar 18, 2013
- Termination Date
- Mar 7, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- List Number 13086: 286 pumps; List Number 13087: 45,376 pumps; List Number 13088: 23,492 pumps
Product Description
Reason for Recall
Distribution Pattern
Worldwide Distribution - USA (nationwide) Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.