Z-1171-2024 Class II Ongoing

Recalled by FUJIFILM Healthcare Americas Corporation — Lexington, MA

Recall Details

Product Type
Devices
Report Date
February 28, 2024
Initiation Date
January 3, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13

Product Description

Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.

Reason for Recall

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Distribution Pattern

US domestic distribution nationwide. International distribution to Israel.

Code Information

UDI-DI 854904006008 Software versions 7.3.000, 7.2.200, 7.2.100, 7.2.000, 7.1.000, 7.0.000