Z-1178-2014 Class II Terminated
Recalled by Lin-Zhi International Inc
— Sunnyvale, CA
FDA device recall Z-1178-2014 was initiated by Lin-Zhi International Inc on Feb 6, 2014 and is designated Class II. Reason for recall: Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards the end of their… The recall status is terminated (terminated Mar 10, 2015). Affected quantity: 130 items.
Recall details
- Product Type
- Devices
- Report Date
- Mar 19, 2014
- Initiation Date
- Feb 6, 2014
- Termination Date
- Mar 10, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 130 items
Product description
Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially infectious. IVD; The Cannabinoids (cTHC also known as THC version c) DAU negative and positive controls are human urine-based liquids and ready to use. The constituent is a processed drug-free human urine matrix, containing buffers, stabilizers, and less than 0.09 % of sodium azide. The controls are prepared by spiking known concentrations of 11-nor-9-THC-9-COOH into the drug-free matrix. cTHC is referenced as an abbreviation for 11-nor-9-THC-9-COOH and it may be referred to simply as THC in the text of this report. The controls contain 37.5 ng/mL and 62.5 ng/mL of cTHC,¿ 10% by GC/MS or LC/MS analysis. Lin Zhi International inc. Sunnyvale, CA 94085 The Cannabinoid Drug of Abuse (DAU) negative and positive controls are used as assay quality control material to monitor the precision of the Lin-Zhi International, Inc. (LZI) Cannabinoid Enzyme Immunoassay (Ref # 0070c) on a number of automated clinical chemistry analyzers.
Reason for recall
Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards the end of their shelf life due to use of dated but unexpired secondary stock used to create the products.
Distribution pattern
US Distribution including the states of SC, NC, LA, FL, TX,, AL and GA and Internationally to Germany.
Code information
Catalog number 0007Cm; lot number 1211158R; Expiration May 14, 2014.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1178-2014
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Lin-Zhi International Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works