Z-1270-2014 Class III Terminated

Recalled by Lin-Zhi International Inc — Sunnyvale, CA

Recall Details

Product Type
Devices
Report Date
April 2, 2014
Initiation Date
February 18, 2014
Termination Date
April 25, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
120 units, 60 of each lot number

Product Description

Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer

Reason for Recall

Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).

Distribution Pattern

US Distribution including CA.

Code Information

Product catalog number: A68827; Lots 1308006, 1308131, Expiry: March 28, 2015.