Z-1270-2014 Class III Terminated

Recalled by Lin-Zhi International Inc — Sunnyvale, CA

FDA device recall Z-1270-2014 was initiated by Lin-Zhi International Inc on Feb 18, 2014 and is designated Class III. Reason for recall: Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP… The recall status is terminated (terminated Apr 25, 2014). Affected quantity: 120 units, 60 of each lot number.

Recall details

Product Type
Devices
Report Date
Apr 2, 2014
Initiation Date
Feb 18, 2014
Termination Date
Apr 25, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
120 units, 60 of each lot number

Product description

Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer

Reason for recall

Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).

Distribution pattern

US Distribution including CA.

Code information

Product catalog number: A68827; Lots 1308006, 1308131, Expiry: March 28, 2015.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1270-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Lin-Zhi International Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1270-2014. Report a data issue: hello@regdatalab.com.