Z-1256-2019 Class III Terminated

Recalled by Lin-Zhi International Inc — Santa Clara, CA

Recall Details

Product Type
Devices
Report Date
May 8, 2019
Initiation Date
January 11, 2019
Termination Date
March 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
49 kits

Product Description

Ethyl Alcohol Enzymatic Assay, (a) REF 0220 (small test kit), (b) REF 0221 (large test kit) Product Usage: Detergent causes faster degradation in EtOH assay. Possible container or pitchers with some detergent residue was used while producing R2 of the defective lot.

Reason for Recall

Shelf life of the product may be reduced due to degradation of the assay.

Distribution Pattern

US Nationwide Distribution in the states of CA, MA, GA, KY, NC, MI, MD, NC, CT, AZ SC, TN, OR

Code Information

(a) REF 0220 (small test kit), Lot Code 1810026 (b) REF 0221 (large test kit), Lot Code 1810025