Z-1184-2021 Class II Ongoing

Recalled by Medical Components, Inc Dba MedComp — Harleysville, PA

FDA device recall Z-1184-2021 was initiated by Medical Components, Inc Dba MedComp on Jan 13, 2021 and is designated Class II. Reason for recall: The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem… The recall status is ongoing. Affected quantity: 96 pieces total (updated 3/11/2021 additional 72 pieces).

Recall details

Product Type
Devices
Report Date
Mar 17, 2021
Initiation Date
Jan 13, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
96 pieces total (updated 3/11/2021 additional 72 pieces)

Product description

Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.

Reason for recall

The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.

Distribution pattern

US Nationwide distribution in the states of CA, DC, GA, IL, IN, MA, MI, NJ, NM, NY, OH, PA, TX. Updated 3/11/2021

Code information

Catalog Number: MRDP50AXS, Lot Number: MPDX180, UDI Number: 884908149753 (updated 3/11/2021 added lot number: MPCY660)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1184-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medical Components, Inc Dba MedComp? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medical Components, Inc Dba MedComp: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medical Components, Inc Dba MedComp →

Source: openFDA record for Z-1184-2021. Report a data issue: hello@regdatalab.com.