Z-2217-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 14, 2019
- Initiation Date
- March 13, 2019
- Termination Date
- March 5, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6830
Product Description
14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425 Product Usage: An aid for introducing a hemodialysis catheter
Reason for Recall
Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
Distribution Pattern
US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
Code Information
MCAM240 MCBL480 MCBS870 MCCF480 MCCJ320 MCCW690 MCCZ500 MCFG830 MCFR160 MCFS760 MCFZ360 MKAJ330 MKAR130 MKBL500 MKBX060 MMAB730