Z-2934-2018 Class II Terminated

Recalled by Medical Components, Inc dba MedComp — Harleysville, PA

Recall Details

Product Type
Devices
Report Date
September 5, 2018
Initiation Date
July 12, 2018
Termination Date
July 2, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Duo-Flow Double Lumen Soft Tip CRRT/ABP Catheter Set; DL11/24 . Hemodialysis catheter insertion kits

Reason for Recall

The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.

Distribution Pattern

Nationwide distribution; worldwide distribution.

Code Information

MCAL840 MCBB800 MCBW450 MCCH690