Z-1227-2024 Class II Ongoing
FDA device recall Z-1227-2024 was initiated by Ortho-Clinical Diagnostics, Inc. on Jan 26, 2024 and is designated Class II. Reason for recall: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient… The recall status is ongoing. Affected quantity: 7 US; 4 OUS.
Recall Details
- Product Type
- Devices
- Report Date
- Mar 13, 2024
- Initiation Date
- Jan 26, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 US; 4 OUS
Product Description
VITROS XT 3400 Chemistry System, Catalog No. 6844458
Reason for Recall
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Code Information
VITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940