Z-1227-2024 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, Inc. — Rochester, NY

FDA device recall Z-1227-2024 was initiated by Ortho-Clinical Diagnostics, Inc. on Jan 26, 2024 and is designated Class II. Reason for recall: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient… The recall status is ongoing. Affected quantity: 7 US; 4 OUS.

Recall Details

Product Type
Devices
Report Date
Mar 13, 2024
Initiation Date
Jan 26, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 US; 4 OUS

Product Description

VITROS XT 3400 Chemistry System, Catalog No. 6844458

Reason for Recall

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

Code Information

VITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1227-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.