Z-1230-2024 Class II Ongoing
FDA device recall Z-1230-2024 was initiated by Ortho-Clinical Diagnostics, Inc. on Jan 26, 2024 and is designated Class II. Reason for recall: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined… The recall status is ongoing. Affected quantity: 15 US; 43 OUS.
Recall Details
- Product Type
- Devices
- Report Date
- Mar 13, 2024
- Initiation Date
- Jan 26, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 US; 43 OUS
Product Description
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Reason for Recall
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Code Information
VITROS System Software Versions 3.8.0 or 3.8.1 Regular UDI-DI 10758750012343 Refurbished UD-DI 10758750033201 Serial Numbers: US: J46000208 J46000209 J46000240 J46000497 J46000864 J46000865 J46000898 J46001030 J46001064 J46001125 J46001132 J46001293 J46001462 J46001703 J46001704 OUS: J46000150 J46000933 J46000162 J46001029 J46000190 J46001041 J46000247 J46001042 J46000296 J46001087 J46000393 J46001119 J46000410 J46001186 J46000478 J46001187 J46000499 J46001210 J46000531 J46001217 J46000564 J46001289 J46000617 J46001541 J46000772 J46001565 J46000788 J46001571 J46000800 J46001609 J46000801 J46001629 J46000812 J46001639 J46000821 J46001645 J46000855 J46001680 J46000859 J46000948 J46000861 J46000983 J46000910