Z-1233-2023 Class II Ongoing

Recalled by Elekta Inc — Saint Charles, MO

FDA device recall Z-1233-2023 was initiated by Elekta Inc on Feb 28, 2023 and is designated Class II. Reason for recall: Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation. The recall status is ongoing. Affected quantity: 2,020 units.

Recall details

Product Type
Devices
Report Date
Mar 15, 2023
Initiation Date
Feb 28, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,020 units

Product description

Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Reason for recall

Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.

Distribution pattern

Nationwide distribution in the US. International distribution to Albania, Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Malta, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Occupied, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Zimbabwe.

Code information

Software Build: 5.11.00, 5.11.01, 5.11.02, 5.11.03; UDI Numbers: (01)00858164002169(10)5.11.00, (01)00858164002169(10)5.11.01, (01)00858164002169(10)5.11.02, (01)00858164002169(10)5.11.03

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1233-2023
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1233-2023. Report a data issue: hello@regdatalab.com.