Z-1241-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 4, 2018
- Initiation Date
- January 30, 2018
- Termination Date
- May 30, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 387 units
Product Description
AltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL TRIAL, STANDARD, REF 804-06-152, Qty: 01, Encore Medical, LP Orthopedic surgical instrument. Product Usage: The humeral socket shell trial is used to replicate/trial the shell geometry of the AltiVate Reverse Standard Shell Humeral Stem implants. It assembles to the humeral broaches by a screw in order to replicate/trial the entire humeral stem implant geometry prior to implantation.
Reason for Recall
Broken or missing screws and retaining clips. This may be due to screw tolerance and material strength of the trial.
Distribution Pattern
US Nationwide Distribution
Code Information
Lot Codes: 183312L03, 200006L01, 224643L01