Z-1241-2018 Class II Terminated

Recalled by Encore Medical, Lp — Austin, TX

Recall Details

Product Type
Devices
Report Date
April 4, 2018
Initiation Date
January 30, 2018
Termination Date
May 30, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
387 units

Product Description

AltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL TRIAL, STANDARD, REF 804-06-152, Qty: 01, Encore Medical, LP Orthopedic surgical instrument. Product Usage: The humeral socket shell trial is used to replicate/trial the shell geometry of the AltiVate Reverse Standard Shell Humeral Stem implants. It assembles to the humeral broaches by a screw in order to replicate/trial the entire humeral stem implant geometry prior to implantation.

Reason for Recall

Broken or missing screws and retaining clips. This may be due to screw tolerance and material strength of the trial.

Distribution Pattern

US Nationwide Distribution

Code Information

Lot Codes: 183312L03, 200006L01, 224643L01