Z-1245-2015 Class II Terminated

Recalled by Baxter Healthcare Corp. — Deerfield, IL

FDA device recall Z-1245-2015 was initiated by Baxter Healthcare Corp. on Apr 3, 2014 and is designated Class II. Reason for recall: Additional warning in the AMIA APD System labeling for patients vulnerable to unintended Increased Intraperitoneal Volume (IIPV), also known as overfill. The patients include neonates, infants, small… The recall status is terminated (terminated Jul 21, 2017). Affected quantity: 55 systems.

Recall Details

Product Type
Devices
Report Date
Mar 11, 2015
Initiation Date
Apr 3, 2014
Termination Date
Jul 21, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
55 systems

Product Description

AMIA v1.0 APD System. For Automated Peritoneal Dialysis (APD) therapy for pediatric and adult renal patients.

Reason for Recall

Additional warning in the AMIA APD System labeling for patients vulnerable to unintended Increased Intraperitoneal Volume (IIPV), also known as overfill. The patients include neonates, infants, small children, and any patient with significant heart and/or lung disease.

Distribution Pattern

US: Nationwide in the states of (GA, MA, NC, NY, PA, WA)

Code Information

Product Code: 5C9310

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1245-2015
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.