Z-1253-2016 Class II Terminated

Recalled by Biomet Spine, LLC — Broomfield, CO

FDA device recall Z-1253-2016 was initiated by Biomet Spine, LLC on Feb 12, 2016 and is designated Class II. Reason for recall: Zimmer Biomet Spine initiated a recall of certain Solitaire Ti Spacers because they may have a mislabeled lordotic angle of 12degrees when the angle is actually 6 degrees. The recall status is terminated (terminated May 10, 2016). Affected quantity: 10.

Recall details

Product Type
Devices
Report Date
Apr 6, 2016
Initiation Date
Feb 12, 2016
Termination Date
May 10, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10

Product description

10mm X 12o Solitaire Ti Medium Spacer  Solitaire Anterior Spinal System, Model Number 1400-1230. Product Usage: The SOLITAIRETM SPINAL SYSTEM is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Additionally, the SOLITAIRETM ANTERIOR SPINAL SYSTEM is indicated for use in the thoracolumbar spine (i.e., T10 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The SolitaireTM System is also indicated for treating fractures of the thoracic and lumbar spine. The SolitaireTM System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.

Reason for recall

Zimmer Biomet Spine initiated a recall of certain Solitaire Ti Spacers because they may have a mislabeled lordotic angle of 12degrees when the angle is actually 6 degrees.

Distribution pattern

Worldwide Distribution - US Nationwide - one foreign consignee (not Canada).

Code information

P/N1400-1230 from Lot#2350871 from either loose inventory or Kit 555000162, Date of Manufacture: April 16, 2013

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1253-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Biomet Spine, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1253-2016. Report a data issue: hello@regdatalab.com.