Z-0983-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- January 21, 2015
- Initiation Date
- August 22, 2013
- Termination Date
- March 13, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14 units
Product Description
Timberline Anchored Lateral Fixed Retractable Sleeve Model No. 8630-0205. Timberline Anchored Lateral Fixed Retractable Sleeves are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae. The Timberline Retractable Drill and the Timberline Retractable Awl fit within this Sleeve. Spinal Surgical Device.
Reason for Recall
The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve.
Distribution Pattern
US Distribution to states of: AZ, CA, NY, TN and TX including PR.
Code Information
LOT Numbers: TU00053A, TU00053B and TU00118 Product Code: 86300205