Z-0983-2015 Class II Terminated

Recalled by Biomet Spine, LLC — Broomfield, CO

Recall Details

Product Type
Devices
Report Date
January 21, 2015
Initiation Date
August 22, 2013
Termination Date
March 13, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 units

Product Description

Timberline Anchored Lateral Fixed Retractable Sleeve Model No. 8630-0205. Timberline Anchored Lateral Fixed Retractable Sleeves are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae. The Timberline Retractable Drill and the Timberline Retractable Awl fit within this Sleeve. Spinal Surgical Device.

Reason for Recall

The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve.

Distribution Pattern

US Distribution to states of: AZ, CA, NY, TN and TX including PR.

Code Information

LOT Numbers: TU00053A, TU00053B and TU00118 Product Code: 86300205