Z-0982-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- January 21, 2015
- Initiation Date
- August 22, 2013
- Termination Date
- March 13, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15
Product Description
Anchored Lateral Retractable Drill, Timberline IPF IFU IFU8621-0113, Model 8630-0102. Timberline Anchored Lateral Retractable Drills are used with the Timberline MPF System to establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.
Reason for Recall
The outer diameter of the Drill shaft is oversized; thereby, resulting in interference fit when inserting the Drill into the Fixed or Variable Sleeve Assembly.
Distribution Pattern
Distributed in PR and the states of NY, AZ, TN, TX, and CA.
Code Information