Z-1256-2018 Class II

Recalled by Encore Medical, Lp — Austin, TX

Recall Details

Product Type
Devices
Report Date
April 4, 2018
Initiation Date
January 29, 2018
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
92 units

Product Description

Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 38mm, REF 804-07-380 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.

Reason for Recall

It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants

Distribution Pattern

US Nationwide Distribution

Code Information

All Serial Numbers