Z-1263-2022 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1263-2022 was initiated by Olympus Corporation Of The Americas on Apr 20, 2022 and is designated Class II. Reason for recall: Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and… The recall status is ongoing. Affected quantity: 2 units.

Recall Details

Product Type
Devices
Report Date
Jun 22, 2022
Initiation Date
Apr 20, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6

Reason for Recall

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

Distribution Pattern

CA FL MA MD MI NJ OH PR WI

Code Information

UDI-DI: 04953170340802 Serial Number: 2836210 2834747I

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1263-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.