Z-1263-2022 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

Recall Details

Product Type
Devices
Report Date
June 22, 2022
Initiation Date
April 20, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6

Reason for Recall

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

Distribution Pattern

CA FL MA MD MI NJ OH PR WI

Code Information

UDI-DI: 04953170340802 Serial Number: 2836210 2834747I