Z-1274-2022 Class II Terminated

Recalled by Olympus Corporation of the Americas — Center Valley, PA

Recall Details

Product Type
Devices
Report Date
June 29, 2022
Initiation Date
April 22, 2022
Termination Date
March 13, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

Colonovideoscope Model CF-H180AL

Reason for Recall

A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding

Distribution Pattern

US Nationwide distribution in the state of KY.

Code Information

UDI-DI: 04953170339615 Serial number: 2501131