Z-1275-2018 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm

FDA device recall Z-1275-2018 was initiated by RAYSEARCH LABORATORIES AB on Jan 26, 2018 and is designated Class II. Reason for recall: Software issue with Center Beam in Field functionality. Issue can result in incorrect treatment volume delivered to patient. The recall status is terminated (terminated Jan 13, 2020). Affected quantity: 2484.

Recall Details

Product Type
Devices
Report Date
Apr 4, 2018
Initiation Date
Jan 26, 2018
Termination Date
Jan 13, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2484

Product Description

Ray Station 4.9 Ray Station 5, Ray Station 6 and Ray Station 7 Software build numbers: 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7 or 7.0.0.19 UDI: 0735000201006820171130

Reason for Recall

Software issue with Center Beam in Field functionality. Issue can result in incorrect treatment volume delivered to patient.

Distribution Pattern

Nationwide

Code Information

Software build numbers: 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7 or 7.0.0.19

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1275-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.