Z-1299-2022 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm

Recall Details

Product Type
Devices
Report Date
July 6, 2022
Initiation Date
June 10, 2022
Termination Date
June 11, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
26

Product Description

RayStation 11B

Reason for Recall

This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.

Distribution Pattern

US Nationwide distribution in the states of NJ and TX.

Code Information

UDI-DI 0735000201042620211208 Software version numbers 12.0.0.932, 12.1.0.1221, 12.0.3.68