Z-1311-2022 Class II Terminated

Recalled by Fresenius Medical Care Holdings, Inc. — Waltham, MA

FDA device recall Z-1311-2022 was initiated by Fresenius Medical Care Holdings, Inc. on Jun 1, 2022 and is designated Class II. Reason for recall: The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of… The recall status is terminated (terminated Oct 7, 2024). Affected quantity: 3 cases.

Recall Details

Product Type
Devices
Report Date
Jul 13, 2022
Initiation Date
Jun 1, 2022
Termination Date
Oct 7, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 cases

Product Description

NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3

Reason for Recall

The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of potential bloodstream infection.

Distribution Pattern

US Nationwide distribution in the states of Texas and Louisiana.

Code Information

UDI: (01)00840861101702 (17) 230131 (10) 21AGAC009 Lot 21AGAC009

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1311-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.