Z-1323-2026 Class I Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
February 18, 2026
Initiation Date
December 22, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
650

Product Description

Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554¿

Reason for Recall

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Distribution Pattern

Distribution US nationwide.

Code Information

UDI-DI (case) 20197344019554¿ (ea) 10197344019557 Lots EP250324 EP250416 EP250512 EP250521 EP250609 EP250630 EP250710 EP250724