Z-1362-2018 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-1362-2018 was initiated by Beckman Coulter Inc. on Mar 2, 2018 and is designated Class II. Reason for recall: Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but… The recall status is terminated (terminated Jul 6, 2023). Affected quantity: 5247 units.

Recall Details

Product Type
Devices
Report Date
Apr 18, 2018
Initiation Date
Mar 2, 2018
Termination Date
Jul 6, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5247 units

Product Description

BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370

Reason for Recall

Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.

Distribution Pattern

Worldwide

Code Information

All

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1362-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.