Z-1362-2018 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

Recall Details

Product Type
Devices
Report Date
April 18, 2018
Initiation Date
March 2, 2018
Termination Date
July 6, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5247 units

Product Description

BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370

Reason for Recall

Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.

Distribution Pattern

Worldwide

Code Information

All