Z-1363-2018 Class II Terminated

Recalled by Orthofix, Inc — Lewisville, TX

Recall Details

Product Type
Devices
Report Date
April 18, 2018
Initiation Date
February 23, 2018
Termination Date
June 18, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
397 units

Product Description

ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003

Reason for Recall

Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.

Distribution Pattern

US

Code Information

UDI 18257200125371, Lot Numbers: O02, O03 and O04