Z-1367-2022 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, Inc. — Rochester, NY

Recall Details

Product Type
Devices
Report Date
July 20, 2022
Initiation Date
May 26, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20858

Product Description

Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems. Product Code: 6199960

Reason for Recall

On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Colombia, India, Singapore.

Code Information

UDI-DI :10758750034413. All expired, in-date and future released lots