Z-1369-2024 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

Recall Details

Product Type
Devices
Report Date
April 3, 2024
Initiation Date
February 14, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
666 units

Product Description

Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on automated analysers Catalog Number: CVS 2699

Reason for Recall

Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods

Distribution Pattern

US Nationwide distribution in the states of CA, IN.

Code Information

GTIN: 05055273207439 (1) Batch 600957 Lot 1378CY - 1382CY (2) Batch 621809, 621815 Lot 1403CY - 1407CY (3) Batch 650533 Lot 1414CY - 1418CY