Z-1593-2024 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North)

FDA device recall Z-1593-2024 was initiated by Randox Laboratories Ltd. on Jan 31, 2024 and is designated Class II. Reason for recall: Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and… The recall status is ongoing. Affected quantity: 8 units.

Recall details

Product Type
Devices
Report Date
Apr 24, 2024
Initiation Date
Jan 31, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 units

Product description

Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567

Reason for recall

Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.

Distribution pattern

US Nationwide distribution in the states of GA, MA, WV including PR.

Code information

GTIN: 05055273204520 (1) Batch/Lot: 619282 / 1094MA-1098MA