Z-1593-2024 Class II Ongoing
FDA device recall Z-1593-2024 was initiated by Randox Laboratories Ltd. on Jan 31, 2024 and is designated Class II. Reason for recall: Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and… The recall status is ongoing. Affected quantity: 8 units.
Recall details
- Product Type
- Devices
- Report Date
- Apr 24, 2024
- Initiation Date
- Jan 31, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8 units
Product description
Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Reason for recall
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
Distribution pattern
US Nationwide distribution in the states of GA, MA, WV including PR.
Code information
GTIN: 05055273204520 (1) Batch/Lot: 619282 / 1094MA-1098MA