Z-1594-2024 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North)

Recall Details

Product Type
Devices
Report Date
April 24, 2024
Initiation Date
January 31, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product Description

Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426

Reason for Recall

Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.

Distribution Pattern

US Nationwide distribution in the states of GA, MA, WV including PR.

Code Information

GTIN: 05055273204544 (1) Batch/Lot: 602114 /1083MA-1087MA (2) Batch/Lot: 636645 /1108MA-1112MA (3) Batch/Lot: 662429 /1083MA- 1087MA