Z-1373-2024 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics — Rochester, NY

FDA device recall Z-1373-2024 was initiated by Ortho-Clinical Diagnostics on Feb 27, 2024 and is designated Class II. Reason for recall: The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could… The recall status is ongoing. Affected quantity: 11,224 units.

Recall Details

Product Type
Devices
Report Date
Apr 3, 2024
Initiation Date
Feb 27, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11,224 units

Product Description

VITROS Performance Verifier II

Reason for Recall

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Code Information

Product Code: 8231474; GTIN: 10758750004577; Lot No. R1176.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1373-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.