Z-1421-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 27, 2022
- Initiation Date
- June 13, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 403 units
Product Description
LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055
Reason for Recall
There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.
Distribution Pattern
Domestic: USA AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TX, UT, VT, WA, WI, International: Belgium, Canada, Croatia, Netherlands, Portugal, Spain
Code Information