Z-3140-2026 Class I Ongoing

Recalled by CooperSurgical, Inc. — Trumbull, CT

FDA device recall Z-3140-2026 was initiated by CooperSurgical, Inc. on Aug 11, 2026 and is designated Class I. Reason for recall: Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectivene… The recall status is ongoing. Affected quantity: 149 boxes (745 units).

Recall Details

Product Type
Devices
Report Date
Sep 30, 2026
Initiation Date
Aug 11, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
149 boxes (745 units)

Product Description

INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.

Reason for Recall

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.

Distribution Pattern

Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.

Code Information