Z-0691-2025 Class II Ongoing
Recalled by CooperSurgical, Inc.
— Trumbull, CT
FDA device recall Z-0691-2025 was initiated by CooperSurgical, Inc. on Nov 15, 2024 and is designated Class II. Reason for recall: The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately. The recall status is ongoing. Affected quantity: 43 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 25, 2024
- Initiation Date
- Nov 15, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 43 units
Product description
Brand Name: K-Systems Product Name: R65 Trolley Model/Catalog Number: K24012 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A
Reason for recall
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
Distribution pattern
States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines. Countries: AT, AU, BE, BH, BR, CA, CH, CN, CZ, DE, EG, ES, ET, FR, GB, GH, HU, ID, IE, IL, IN, IQ, IT, JO, JP, KR, KW, LB, MA, MK, MT, NL, NO, PE, PH, PK, PT, QA, RO, RS, SA, SD, SE, SO,TR, TW, TZ, US, ZA
Code information
Version or Model: K24012; Primary DI Number: 00888937025026; Serial Numbers: 2234KH01, 2234KH02, 2234KH03, 2234KH04, 2234KH05, 2234KH06, 2234KH07, 2234KH08, 2234KH09, 2234KH10, 2235KH41, 2235KH42, 2235KH43, 2235KH44, 2235KH45, 2235KH46, 2235KH47, 2235KH48, 2235KH49, 2235KH50, 2235KH51, 2235KH52, 2235KH53, 2235KH54, 2235KH55, 2235KH56, 2235KH57, 2235KH58, 2235KH59, 2235KH60, 2243KH01, 2243KH02, 2243KH03, 2243KH04, 2243KH05, 2243KH06, 2243KH07, 2243KH08, 2243KH09, 2243KH10, 2243KH11, 2243KH12, 2243KH13, 2243KH14, 2243KH15, 2243KH16, 2243KH17, 2243KH18, 2243KH19, 2243KH20, 2307KH01, 2307KH02, 2307KH03, 2307KH04, 2307KH05, 2307KH06, 2307KH07, 2307KH08, 2307KH09, 2307KH10, 2320KH11, 2320KH12, 2320KH13, 2320KH14, 2320KH15, 2320KH16, 2320KH17, 2320KH18, 2320KH19, 2320KH20, 2337KH11, 2337KH12, 2337KH13, 2337KH14, 2337KH15, 2337KH16, 2337KH17, 2337KH18, 2337KH19, 2337KH20, 2340KH01, 2340KH02, 2340KH03, 2340KH04, 2340KH05,
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0691-2025
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for CooperSurgical, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works