Z-1427-2022 Class II Ongoing

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-1427-2022 was initiated by GE Healthcare, LLC on May 27, 2022 and is designated Class II. Reason for recall: The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a… The recall status is ongoing. Affected quantity: 667 devices.

Recall Details

Product Type
Devices
Report Date
Jul 27, 2022
Initiation Date
May 27, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
667 devices

Product Description

GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, Part # 1011-9050-000.

Reason for Recall

The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person.

Distribution Pattern

Distribution was made to AZ, CA, CO, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MT, NC, NE, NV, NY, OH, PA, SC, TN, TX, VA, and WV. There was also military distribution but no government distribution. Foreign distribution was made to Australia, Canada, China, Estonia, Finland, France, Germany, Greece, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Malaysia, New Zealand, Philippines, Romania, Russia, Saudi Arabia, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

Code Information

Avance CS2 and Avance CS2 Pro - GTIN 00840682102322, all serial numbers; Aisys CS2 - GTIN 00840682102292, all serial numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1427-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.