Z-1451-2015 Class II Terminated

Recalled by Baxter Healthcare Corp — Deerfield, IL

FDA device recall Z-1451-2015 was initiated by Baxter Healthcare Corp on Mar 2, 2015 and is designated Class II. Reason for recall: Loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in… The recall status is terminated (terminated May 22, 2018). Affected quantity: 58 MDLs.

Recall Details

Product Type
Devices
Report Date
Apr 22, 2015
Initiation Date
Mar 2, 2015
Termination Date
May 22, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
58 MDLs

Product Description

Master Drug Library Software version 8.0, Product Code 35723V080, to be used with SIGMA Spectrum Infusion System (Pump) version 8.0, Product Code 35700BAX2 The SIGMA Spectrum Infusion Pump with Master Drug Library (MDL) is intended to be used for the controlled administration of fluids.

Reason for Recall

Loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in minutes on the pump dose setup screen (20 sec may show as 1 min on the pump display). The pump will administer drugs as configured. The discrepancy may cause therapy delay or unintended rate of delivery.

Distribution Pattern

Worldwide Distribution: US Distribution to states of: CA, CT, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, NC, NE, NH, NY, SD, TX, UT, WV, including Puerto Rico and country of: Canada.

Code Information

Software version 8.0, Product Code 35723V080

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1451-2015
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.