Z-1468-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-1468-2026 was initiated by Medline Industries, LP on Jan 8, 2026 and is designated Class II. Reason for recall: Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid… The recall status is ongoing. Affected quantity: 150.

Recall details

Product Type
Devices
Report Date
Mar 4, 2026
Initiation Date
Jan 8, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
150

Product description

Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

Reason for recall

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Distribution pattern

US Nationwide distribution.

Code information

UDI-DI (ea) 10195327015954 UDI-DI (case) 40195327015955 Kit lots 24IME953 25CMA708

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1468-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1468-2026. Report a data issue: hello@regdatalab.com.