Z-1468-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- March 4, 2026
- Initiation Date
- January 8, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 150
Product Description
Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167
Reason for Recall
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Distribution Pattern
US Nationwide distribution.
Code Information
UDI-DI (ea) 10195327015954 UDI-DI (case) 40195327015955 Kit lots 24IME953 25CMA708